Smart cooled incubator

Introducing the new generation of FG smart cooled incubators

Cooling system with unique NON-CUT technology

Ability to work in high temperature environments (up to 30°C)

Temperature fluctuation ±0.1 degree even in cooling mode

40% energy saving

No thrush

The most accurate chamber in the world with unique isotherm in accordance with the most reliable international standards

Quick temperature return to the set temperature after opening and closing the device door

Better Uniformity and faster heating with the second-generation HCT

Uniformity @ 27 points of the chamber

Temperature stability

dual chamber

Double safety to protect your tests against unwanted increase or decrease in temperature

Adjustable safety thermostat with separate and independent sensor

According to the DIN 12880 Class 3.3 standard

Temperature control in case of unwanted increase

Two PT100 sensors

Image
Product video

To get to know More details of the product Watch the opposite video.

Product video gallery

Table of technical specifications of CIF model laboratory refrigerated incubator
Image link
External specifications
CIF 55S
CIF 120S
CIF 240S
CIF 700S
Table of technical specifications of CIF model laboratory refrigerated incubator
16
The minimum distance allowed from the back wall (cm)
10
10
10
20
Depth (with handle and controller) (G) (cm)
60
73
74
90
Depth (without handle) (F) (cm)
54.5
67.5
69
85
Height of the device with base (E) (cm)
84.5
89
148
191
Device width (D) (cm)
64.5
82.5
90
121
Device weight (kg)
72
105
110
325
number of doors
1
1
1
2
Image link
Internal specifications
CIF 55S
CIF 120S
CIF 240S
CIF 700S
Internal volume (LIT)
50
Height (B) (cm)
41.5
46
80.5
128
Width (A) (cm)
41
59
66.5
97
Depth (C) (cm)
29.5
42
42.5
57
Weight capacity for each tray (kg)
30
30
40
40
Ability to bear the weight of the entire floor (kg)
60
100
150
250
The number of standard trays in the machine - (can be ordered)
2 – (4)
2 – (5)
2 – (8)
2 – (14)
circulating fan
have
have
have
have
Heat information
CIF 55S
CIF 120S
CIF 240S
CIF 700S
Thermal range:
5- to 80
Display Accuracy at 5°C (Deviation: ±K)
±0.1
±0.1
±0.1
±0.1
Display Accuracy at 37°C (Deviation: ±K)
±0.1
±0.1
±0.1
±0.1
Display Accuracy at 70°C (Deviation: ±K)
±0.1
±0.1
±0.1
±0.1
Temperature Fluctuation at 5°C: ±K
±0.2
±0.1
±0.2
±0.5
Temperature Fluctuation at 37°C: ±K
±0.2
±0.1
±0.5
±0.5
Temperature Fluctuation at 70°C: ±K
±0.2
±0.1
±0.5
±0.6
Temperature Control Accuracy
±0.1
±0.1
±0.1
±0.1
Uniformity at 27 points
CIF 55S
CIF 120S
CIF 240S
CIF 700S
Temperature Uniformity at 5°C: ±K
±0.9
±0.7
±0.9
±1.0
Temperature Uniformity at 37°C: ±K
±0.4
±0.3
±0.8
±0.8
Temperature Uniformity at 70°C: ±K
±0.9
±0.6
±1.4
±1.5
heating-up time
CIF 55S
CIF 120S
CIF 240S
CIF 700S
Time to Reach 5°C (min)
45
Time to Reach 37°C (min)
20
20
15
30
Time to Reach 70°C (min)
25
65
40
60
Returning the temperature to the Set temperature after opening the device door for 30 seconds (*) (Recovery time)
CIF 55S
CIF 120S
CIF 240S
CIF 700S
Temperature Recovery Time at 5°C (min)
5
15
12
85
Temperature Recovery Time at 37°C (min)
5
10
5
15
Temperature Recovery Time at 70°C (min)
7
15
10
35
Electrical information
CIF 55S
CIF 120S
CIF 240S
CIF 700S
power consumption (W)
660
Frequency (HZ)
50
50
50
50
Voltage (V AC)
220
220
220
220
flow
3
3.3
7
10
type of electricity
single phase
single phase
single phase
single phase
Image link
Image link
Image link
Image link
Customized equipment and services

Tray made of mesh stainless steel sheet

Manufacturing based on custom dimensions and temperatures

Installation of valves in customized number and diameter

Glass door with metal frame

IQ, OQ, PQ Certificates

Calibration Certificate

Light bulb

net tray

Frequently Asked Questions(FAQ)

It always takes time for the internal chamber temperature to reach the set point. The smart timer will show you the calculated time will be activated when the temperature of the device reaches the set point.

Yes, using a safety and adjustable thermostat in FG devices that acts like a fuse and turns off the refrigerated incubator device if the temperature rises above the set point.

According to FG standards and valid world standards, the best distance from the walls, floor, and ceiling of the incubator is 10% of the length of the same side. And if the depth of the device is 50 cm, 10% of it, that is, 5 cm from the front and back, should be considered the distance from the walls of the device for loading.

Using a safety thermostat in FG devices that works like a fuse and turns off the oven device if the temperature rises above the set point.

Yes, it is possible to create a vent or install a socket for FG incubators upon request.

Yes, you can move the trays if needed and according to the different drawers installed in the cooled incubator.

If you need to make maximum use of the internal space of the device, you can use more trays in the incubator. You can use more space just by buying a tray, and there is no need to pay extra for installing rails or other changes in the device.

ask us

If you still haven’t found the answer to your question, ask us.

"*" indicates required fields

first name and last name*
FG products

To see another FG products. Refer to the products page.

All products
The latest news and blogs

Leave a Reply

Your email address will not be published.Required fields are marked *