Cooled incubator

Cooled incubator

Introducing the New Generation of FG Smart Cooled Incubator

Cooling system with unique NON-CUT technology

Capability to operate in high ambient temperatures (up to 30°C)

Temperature fluctuation of ±0.1°C even in cooling mode

40% energy savings

Frost-free operation

HCT (Homogenic chamber technology)

Rapid return to the set temperature after opening and closing the door of the device

Better thermal equilibrium and faster heating with the second-generation HCT

Temperature uniformity in 27 points of the chamber

Temperature stability

Triple-layer walls

DUAL SAFETY

Adjustable safety thermostat with a separate and independent sensor

Compliant with DIN 12880 Class 3.3 standard

Temperature control in case of unwanted temperature increase

Equipped with two PT100 sensors

Image

Product video

To get to know More details of the product Watch the opposite video.

Product video gallery

Table of technical specifications of CIF model laboratory refrigerated incubator

Exterior dimensions
CIF 55S
CIF 120S
Wall clearance, side (cm)
 
16
Wall clearance, rear (cm)
 
10
10
Depth Plus door handle and controller (cm) (G)
 
60
73
Depth without door handle and controller (cm) (F)
 
54.5
67.5
Height (incl. feet) (cm) (E)
 
84.5
89
Width (cm) (D)
 
64.5
82.5
Weight (empty) (kg)
 
72
105
Number of doors (ea.)
 
1
1
Interior dimensions
CIF 55S
CIF 120S
Interior volume (l)
 
50
Height (cm) (B)
 
41.5
46
Width (cm) (A)
 
41
59
Depth (cm) (C)
 
29.5
42
Load per rack (kg)
 
30
30
Permitted total load (kg)
 
60
100
Racks number standard (max)
 
2 – (4)
2 – (5)
Circulation fan
 
Yes
Yes
Temperature data
CIF 55S
CIF 120S
Temperature range (°C)
 
-5 to 80
Display accuracy (deviation) @ 5˚C (±K)
 
±0.1
±0.1
Display accuracy (deviation) @ 37˚C (±K)
 
±0.1
±0.1
Display accuracy (deviation) @ 70˚C (±K)
 
±0.1
±0.1
Temperature fluctuation @ 5°C (±K)
 
±0.2
±0.1
Temperature fluctuation @ 37°C (±K)
 
±0.2
±0.1
Temperature fluctuation @ 70°C (±K)
 
±0.2
±0.1
Setting temperature accuracy
 
±0.1
±0.1
Uniformity @ 27 points of chamber (variation)
CIF 55S
CIF 120S
@ 5˚C (±k)
 
±0.9
±0.7
@ 37°C (±k)
 
±0.4
±0.3
@ 70°C (±k)
 
±0.9
±0.6
Heating-up/Cooling-down time *
CIF 55S
CIF 120S
Cooling down time to 5°C (min)
 
50
Heating up time to 37°C (min)
 
20
20
Heating up time to 70°C (min)
 
30
65
Recovery time after doors were open for 30 sec *
CIF 55S
CIF 120S
@ 5°C (min)
 
5
@ 37°C (min)
 
5
10
@ 70°C (min)
 
10
15
Electrical data
CIF 55S
CIF 120S
Power (W)
 
660
Frequency (HZ)
 
50
50
Voltage (V AC)
 
220
220
Current(A)
 
3
6
Power type
 
Single phase
Single phase

Customized equipment and services

Perforated stainless steel tray

Custom manufacturing based on specific dimensions and temperatures

Installation of vents in customized quantities and diameters

Glass door with a metal frame

IQ, OQ, PQ certifications

Provision of calibration certificate

Lighting lamp

Wire mesh tray

Frequently Asked Questions(FAQ)

It always takes time for the internal chamber temperature to reach the set point. The smart timer will show you the calculated time will be activated when the temperature of the device reaches the set point.

Yes, using a safety and adjustable thermostat in FG devices that acts like a fuse and turns off the refrigerated incubator device if the temperature rises above the set point.

According to FG standards and valid world standards, the best distance from the walls, floor, and ceiling of the incubator is 10% of the length of the same side. And if the depth of the device is 50 cm, 10% of it, that is, 5 cm from the front and back, should be considered the distance from the walls of the device for loading.

Using a safety thermostat in FG devices that works like a fuse and turns off the oven device if the temperature rises above the set point.

Yes, it is possible to create a vent or install a socket for FG incubators upon request.

Yes, you can move the trays if needed and according to the different drawers installed in the cooled incubator.

If you need to make maximum use of the internal space of the device, you can use more trays in the incubator. You can use more space just by buying a tray, and there is no need to pay extra for installing rails or other changes in the device.

ask us

If you still haven’t found the answer to your question, ask us.

"*" indicates required fields

first name and last name*

Other FG products

To see another FG products. Refer to the products page.

All products

The latest news and blogs

Leave a Reply

Your email address will not be published.Required fields are marked *