Smart laboratory oven

Smart laboratory oven

Introducing the new generation of FG smart laboratory ovens

The most accurate chamber in the world with unique isotherm in accordance with the most reliable international standards

Quick temperature return to the set temperature after opening and closing the device door

Better Uniformity and faster heating with the second-generation HCT

Uniformity @ 27 points of the chamber

Temperature stability

dual chamber

Double safety to protect your tests against unwanted increase or decrease in temperature

Adjustable safety thermostat with separate and independent sensor

Temperature control in case of unwanted temperature increase

According to DIN 12880 Class 3.1 standard

Two PT100 sensors

Video introduction of the laboratory oven

To get to know More details of the product Watch the opposite video.

Product video gallery

Customized equipment and services

Tray made of mesh stainless steel sheet

Installation of valves in customized number and diameter

Glass door with metal frame

IQ, OQ, PQ Certificates

Calibration Certificate

Light bulb

External sensor

net tray

Door lock

FG Smart Laboratory Oven Technical Specifications Table

External specifications
BF55
BF120
BF240
BF400
BM55
Minimum allowed distance from walls and adjacent devices (cm)
16
16
16
16
The minimum distance allowed from the back wall (cm)
10
10
10
10
10
Air vent diameter (cm)
5
5
5
5
5
Depth (with handle and controller) (G)(cm)
60
73
77.2
85.8
60
Depth (without handle) (F)(cm)
54.5
67.5
70.2
84.7
54.5
Height of the device with stand (E)(cm)
63.5
68
105.2
145.5
63.5
Machine width (D)(cm)
64.5
82.5
90
90
54.5
Device weight (kg)
44
62
98
157
41
number of doors
1
1
1
1
1
Internal specifications
BF55
BF120
BF240
BF400
BM55
Internal volume (LIT)
55
250
420
55
Height (B)(cm)
41.5
46
80.5
111.5
41.5
Width (A)(cm)
41
59
66.4
66.4
41
Depth (C) (cm)
32.3
44.5
47
57.1
32.3
Weight capacity for each tray (kg)
30
30
40
40
30
Ability to bear the weight of the entire floor (kg)
60
100
150
200
60
The number of standard trays in the machine - (can be ordered)
2 – (4)
2 – (5)
2 – (8)
2 – (13)
2 – (4)
circulating fan
have
have
have
have
have not
Heat information
BF55
BF120
BF240
BF400
BM55
Thermal range: 15 degrees above ambient temperature (C˚)
80°C
80°C
80°C
80°C
Display accuracy at 60°C (deviation) (±K)
±0.1
±0.1
±0.1
±0.2
Display accuracy at 150°C (deviation) (±K)
±0.2
±0.2
±0.1
±0.1
±0.2
Temperature fluctuation at 60°C (±K)
±0.1
±0.2
±0.2
±0.5
Temperature fluctuation at 150°C (±K)
±0.4
±0.5
±1.0
±0.5
±1.0
Uniformity at 27 points
BF55
BF120
BF240
BF400
BM55
Uniformity at (variation) 60°C (±K)
±1.0
±0.7
±1.5
Uniformity at 150°C (variation) (±K)
±3.0
±3.0
±4.0
±5.0
±3.0
Time to reach temperature*
BF55
BF120
BF240
BF400
BM55
Time to reach temperature 60°C (min)
40
15
-
30
Time to reach temperature 150°C (min)
25
35
25
35
40
Returning the temperature to the Set temperature after opening the device door for 30 seconds (*) (Recovery time)
BF55
BF120
BF240
BF400
BM55
Return temperature in 60°C (min)
8
10
-
25
Return temperature in 150°C (min)
8
10
10
12
20
Ventilation (Air exchange)
BF55
BF120
BF240
BF400
BM55
At 60°C (x/h)
50±5
25±5
-
30±5
At 150°C (x/h)
50±5
25±5
15±5
20±5
10±5
Electrical information
BF55
BF120
BF240
BF400
BM55
power consumption (W)
1320
3520
3762
1320
Frequency (HZ)
50
50
50
50
50
Voltage (V AC)
220
220
220
220
220
flow
6
7.2
16
17.1
6
type of electricity
single phase
single phase
single phase
single phase
single phase

Frequently Asked Questions(FAQ)

According to FG standards and the world’s valid standards, the best distance from the walls, floor and ceiling of the oven is 10% of the length of the same side. For example, if the width of the device is 40 cm, 10% of it means 4 cm on each side (left and right) and if the depth of the device is 50 cm, 10% of it means 5 cm from the front and back, the distance from the walls of the device should be considered for loading.

Yes, with the help of the HCT technology used in the oven, even in the full load mode of the device (according to the principles of loading), you will have full uniformity.

The oven device reaches the set point temperature as quickly as possible and the temperature gradient adjustment is to prolong the time for the chamber temperature to reach the set point temperature, as a result, the time for the chamber temperature to reach the set point temperature cannot be made faster.

Using a safety thermostat in FG devices that works like a fuse and turns off the oven device if the temperature rises above the set point.

It always takes time for the internal chamber temperature to reach the set point. The Smart timer is activated when the temperature reaches the set point.

The cost of changing the color of devices is very high, so it is not cost-effective to do this.

Yes, based on the customer’s order, this feature is available for an additional fee.

The temperature range that is written about ovens, they usually have the ability to control in that range, but they are not very accurate at these temperatures or require changes in the settings of the device. Therefore, for temperatures lower than 80 degrees, usually the best option is an incubator.

No, the mechanism of the air release valve is the maximum exit of vapors and gases created in the chamber. But the circulator fan is used to accelerate the air circulation inside the chamber.

No, if you need more trays in the oven, you can use more space just by buying a tray, and there is no need to install rails or other changes in the device.

Yes, if needed, you can move the trays according to the different drawers installed in the oven.

Yes, a special mechanism has been considered for the air discharge process and the entry of fresh air into the oven chamber, which causes the maximum discharge of moisture and gases created in the chamber.

ask us

If you still haven’t found the answer to your question, ask us.

"*" indicates required fields

first name and last name*

FG products

To see another FG products. Refer to the products page.

All products

The latest news and blogs

Leave a Reply

Your email address will not be published.Required fields are marked *